1. Data Upload
Supports various EPR data formats, including standard text formats and instrument raw data.
- Supported Formats: .txt, .csv, .asc, .dat, .epr, .eprp, .spc, .xlsx
- Bruker Data: Supports uploading
.DSCand.DTAfiles simultaneously (please select both files). - Microwave Frequency: If not included in the file, you will be prompted to enter it manually (common X-band is ~9.8 GHz).
1.1 Text-like Formats
- .txt / .csv / .asc:
- Recommended to have two columns of numbers: first column for magnetic field (G or mT), second for intensity.
- Columns can be separated by spaces, Tabs, or commas. Ensure consistency within the file.
- First row can be headers (e.g.
Field,Intensity) or data; lines starting with#are ignored as comments. - Example (space separated):
# B(G) Intensity3300 0.00233301 0.0028
- .xlsx:
- Uses the first sheet;
- First two columns are Field and Intensity (e.g.,
Col A = Field / Col B = Intensity); - Do not merge cells or insert complex headers in the data region. Extra text above the data is fine.
1.2 Instrument Raw Formats
- .dat / .spc: Directly upload raw spectral files exported from instruments. No formatting needed.
- .epr / .eprp: Software-specific formats containing parameters and data. Upload directly.
- .DSC + .DTA: For Bruker EPR data, select both files to upload together. The system parses them automatically.
2. Analysis Modes
The system provides three analysis modes for different experimental scenarios:
Suitable for pure systems containing only one radical signal (e.g., pure DMPO-OH).
Suitable for mixed systems with two radicals (e.g., DMPO-OH + DMPO-O2-). The system automatically separates them and calculates proportions.
Suitable for complex systems with three radicals. Computationally intensive; use only when three species are confirmed.
3. Manual Peak Picking
New Feature: After loading data, you don't have to wait for AI analysis. Click the 🎯 Manual Peak button next to the delete button to enter manual mode instantly.
- Click 🎯 Manual Peak to enter peak selection mode (button changes to ✓).
- Click on peak positions in the spectrum area; the system will mark them with red dots.
- It is recommended to mark at least 2 main peaks (or all obvious peaks for complex spectra).
- After marking, click 🚀 AI Analysis at the top. The algorithm will force the fit to pass through your marked peaks.
- Click the manual peak button again to exit the mode; click "Clear Peaks" to reset marks.
4. Viewing Results
After analysis, the interface displays comprehensive results:
- Fitted Spectrum: Red dashed line is the fit, black solid line is the experiment. Zoom and pan to inspect details.
- Parameter List: The left table shows the Top 5 matches. Click different rows to view fits for different candidates.
- Multi-component Details: In multi-component mode, click a result to see parameters (g-value, hyperfine coupling A, linewidth) for each component below.
5. AI Report
Click "Generate Detailed AI Report" to let the large language model create a professional analysis report.
- Experimental Conditions: Click "Experiment Settings" before generating to add spin trap and solvent info for a more accurate report.
- Smart Q&A: After report generation, you can continue asking questions about it in the chat box below.
6. Paper Figure
Generate publication-ready figures and captions with one click.
- Format: Generates Nature-style figures by default (high resolution). PNG download supported.
- Caption: Automatically generates English/Chinese figure captions containing experimental conditions and fit parameters. Text download supported.
- WYSIWYG: Generated based on the selected fit result. Click "Download Details" to get full Figure Description Markdown.
7. Typical Example
Example: Analyzing DMPO-OH:
Data Upload
Drag and drop the data file (e.g., sample.epr) into the upload area.
Run Analysis
Click "AI Analysis". Set Spin Trap to "DMPO" and Solvent to "Water/Aqueous".
Check Results
Review the Rank 1 result. If the fitted curve agrees well with the experimental spectrum, prioritize DMPO-OH as a candidate for review alongside experimental conditions and control experiments.
💡 Tip: If auto-fitting is not optimal, try using "Manual Peak" to mark peaks and re-analyze.